D5028 Standard Test Method for Curing Properties of Pultrusion Resins by Thermal Analysis has been revised to D5028-17

Significance and Use

5.1 Differential scanning calorimeters are used to determine chemical reaction thermal profiles of materials. One such reaction is the curing of thermosetting resins.

5.2 This test method is useful for both specification acceptance and for research.

1. Scope

1.1 This test method covers determination of curing parameters of pultrusions resins by differential scanning calorimetry.

1.2 This test method is applicable to pultrusion resin solutions with adequate initiator(s).

1.3 The normal operating temperature range is from 0 to 200°C.

Note 1: Resin systems that do not form an adequate baseline are not covered by this test method.

1.4 Computer or electronic based instruments or data treatment equivalent to this practice are suitable for use.

1.5 The values stated in SI units are to be regarded as standard.

1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific hazard statements, see Note 1.

Note 2: There is no known ISO equivalent to this standard.

1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.


2. Referenced Documents (purchase separately) The documents listed below are referenced within the subject standard but are not provided as part of the standard.

ASTM Standards

D883 Terminology Relating to Plastics

D3418 Test Method for Transition Temperatures and Enthalpies of Fusion and Crystallization of Polymers by Differential Scanning Calorimetry

D3918 Terminology Relating to Reinforced Plastic Pultruded Products

E473 Terminology Relating to Thermal Analysis and Rheology

E967 Test Method for Temperature Calibration of Differential Scanning Calorimeters and Differential Thermal Analyzers

E2160 Test Method for Heat of Reaction of Thermally Reactive Materials by Differential Scanning Calorimetry